47
Troubleshooting
If the equipment fails to operate correctly, the table below suggests
some simple checks and corrective actions. If these steps fail to resolve
the problem, contact Arjo or an approved service agent.
Symptom Possible cause Action
“Beeping” sound when
using the bed
Bed is operating from the
backup battery
Check the power supply cord is
plugged in and the electricity supply
is OK
Check fuse in mains plug (where
fitted)
One or more bed functions
inoperative
Function(s) locked on
ACP
Unlock function(s) on ACP
All functions inoperative
except for raising mattress
platform
Height control software
error
Raise the mattress platform to
maximum height to reset software
Bed is difficult to
manoeuvre around corners
Brake pedals in “steer”
position
Place brake pedals in the “free”
position
All indicators on ACP lit or
flashing
Duty cycle of electrical
system exceeded
Refer to section Duty cycle lockout