EasyManua.ls Logo

Arjo Enterprise 5000X - Fault Indications; Product Lifetime

Arjo Enterprise 5000X
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
48
Fault indications
The bed’s control software indicates problems in the electrical system
by means of flashing indicators on the Attendant Control Panel (ACP).
If you experience any of the indications below, contact Arjo or an
approved service agent.
Indication Possible cause
ACP mattress platform height and head
down tilt indicators flashing
Height actuator fault (foot end)
ACP mattress platform height and foot
down tilt indicators flashing
Height actuator fault (head end)
ACP backrest indicator flashing Backrest actuator fault
ACP thigh section indicator flashing Thigh section actuator fault
Mattress platform height, tilt, backrest and thigh section
indicators flashing.
Control unit fault
Product lifetime
The lifetime of this equipment is typically ten (10) years. “Lifetime” is
defined as the period during which the product will maintain the
specified performance and safety, provided it has been maintained and
operated in conditions of normal use in accordance with the
requirements in these instructions.

Related product manuals