Product Specifications
Test Strips: GLUCOCARD™ 01-SENSOR.
Result Range: 20-600 mg/dL .
Calibration: Plasma referenced.
Sample Size: Minimum 0.3 µL.
Blood Source: Fresh capillary whole blood.
Hematocrit Range: 30% - 54%.
Test Time: 7 seconds.
Assay Method: Biosensor, Glucose Oxidase (Aspergillus niger sourced).
Power Source: One 3-volt lithium battery (CR2032 or DL2032)
Battery Life: Approximately 3,000 tests
Screen Type: Liquid Crystal Display (LCD).
Memory: 50 results with time & date stamp.
Average: 7-day, 14-day or 30-day averaging.
AutomaticShutOff: 3–5minutesafterlastuseraction.
Size: 3.6" x 1.2" x 0.5" (92mm x 30mm x 14 mm).
Weight: 1.1 oz (with battery).
Operating Ranges: 50°F-104°F (10°C-40°C), 20% - 80% relative humidity.
Altitude: 10,000 ft. (3,048 meters).
Unit of Measurement: mg/dL.