EasyManua.ls Logo

Atlas SUNTEST XLS - Calibration; Calibration of the Irradiance with the XenoCal Global-Sensor

Atlas SUNTEST XLS
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating manual SUNTEST XLS / XLS+
- 26 -
1
2
!
Adapter table for XenoCal Global-Sensor
"
Sensor receptacle for the black standard sensor
#
Connector cable
$
XenoCal Global-Sensor
8
Calibration
Figure 22
8.1 Calibration of the irradiance with the
XenoCal Global-Sensor
The equipment should be calibrated before and after eve-
ry test.
The optional XenoCal Global Sensor and an adapter for
the XenoCal sensor (see chapter 5 Accessories) are
necessary for the calibration.
1. Mount the adapter table
!
for the XenoCal Global-
Sensor instead of the sample table in the test cham-
ber.
Mount the adapter table
!
in such a way that the sen-
sor receptacle for the black standard sensor
")
is at
rightfront side of the test chamber.
2. Insert the connector cable
#
in the XenoCal Global
sensor
$
and guide the connector cable
#
below
the adapter table
!
to the PC.
3. Start the program XenoSoft and ONLINE-measure-
ment.
4. In the Suntest XLS+, start the calibration program and
input the irradiance at which the calibration is to be
done.
5 After about 30 minutes, input the actual irradiance va-
lue displayed in the XenoSoft to the corresponding
field in the display unit and confirm with ENTER.
6. At the end of the automatic cooling period, the cali-
bration factor calculated by the software is indicated
in the display unit and automatically taken into account
during every subsequent test.
Test chamber door side

Other manuals for Atlas SUNTEST XLS

Related product manuals