13
8.0 Cleaning / Disinfection
● The complete surface of the hose rewind, the trolley and the
device support must always be cleaned with a damp (not wet)
cloth and disinfected with a surface disinfectant stated
on page 14 prior to use the device in another patient.
● In case the device is used in one patient only the device sur-
face should be cleaned if it is contaminated however at least
once every week with a damp (not wet) cloth and afterwards be
disinfected with a surface disinfectant stated on page 14.
8.2 Oversuction
When is a suction unit oversucked?
A suction device is oversucked if suction material penetrated into
the interior of the device.
How can one realise that the suction device is contaminated?
Generally a reduced suction capacity is a sign of possible oversuction.
The ATMOS C 161 / C 261 Aspirator have a condensate collector at
the bottom of the device. For visual inspection please remove the
cap. The device is oversucked if humidity or contamination is visible
in the condensate collector.
Measures
Thedevicemustbereprocessedbythemanufactureroracertied
ATMOS partner if the device is oversucked or if any reservations exist
regarding the hygiene condition.
A contaminated suction device bears a risk for the patient as well as
for the caregiver. Therefore, we recommend regular checking of the
condensate collector.
8.3 Cleaning instructions
for use in another patient
For use in one
patient
contamination
after suction procedure
1x per day
1x per week
every 2 weeks
every 4 weeks
Exchangebacteriallter X X X
Rinsing the suction hose X X
Disinfecting/autoclaving
the suction hose
X X
Exchanging the suction hose X
Exchangingthengertip X X
Emptying the secretion canister X X
Rinsing the secretion canister X X
Disinfecting the secretion canister X X
Rinsing the lid components X X
Disinfecting the lid components X X
Cleaning of the device surface X X X
Wipe disinfection of the
device surface
X X X
Rinsing the vacuum hose X X
Disinfection of the vacuum hose X X
Disinfection of the hose connector X X
Please observe the instructions for use of the
relevant disinfectant.
Information on how to remove the secretion
canisterlidyouwillndinchapter5.2on
page 9.
● Werecommendtothoroughlyrinsethelidandits
components with clear water after each suction
procedure in case the device is used in one patient
only.
Attention!Pleaseremovebacteriallterpriorto
this!
8.1.6 Device surface
● Priortousethedeviceinanotherpatientthe
complete device surface must always be cleaned
with a damp (not wet) cloth and disinfected with a
surface disinfection solution stated on page 14.
● Incasethedeviceisusedinonepatientonlythe
device surface should be cleaned if it is
contaminated however at least once every week with
a damp (not wet) cloth and afterwards be disinfected
with a surface disinfectant stated on page 14.
Discolouration of the surface could occur by some
disinfectants;afactwhichdoesnottakeeecton
the attributes of the material.
Prior to cleaning please make sure to disconnect the
device from the mains!
The device may never be autoclaved, rinsed under
running water or immersed into any liquids!
8.1.7 Rinsing Canister
● Therinsingcanistermustalwaysbedisinfected
in case the device is used in another patient.
Prior to disinfection please take care to empty
the canister and to rinse it with clear water in
ordertoincreasecleaningeciency.
● Incasethedeviceisusedinonepatientonly
we recommend disinfection of the canister
every week.
● Therinsingcanistermayonlybecleanedwithaph
neutral cleaning liquid which does not contain the
following ingredients:
ammoniac, amines, amides, phenol derivates, ani-
onic tensides.
● Thedisinfectionisexclusivelyallowedwithdisinfect-
ant which does not contain the following ingredients:
alcohol, aromatic hydrocarbons, ammonia, amine.
● Cleaninginadishwasherispossiblewhenusingph
neutral cleaning liquids (5 cycles).
8.1.8 Accessories
● Hoserewind(313.0007.0)/
Trolley (320. 0070.0) /
Device support (313.0012.0)
!
!
!
!
!