17
9.1 Basic information
● Carryoutavisualinspectionoftheunitpriortoeach
use including hoses, collection canister and connec-
tion cable. Damaged cables and hoses must be
replaced immediately.
● Maintenance,repairsandperiodtestsmayonlybe
carried out by persons who have the appropriate
technical knowledge and are familiar with the product.
To carry out these measures the person must have the
necessary test devices and original spare parts.
ATMOS recommends: work should be carried out by
an authorized ATMOS-service partner. This ensures
that repairs and testing are carried out professionally,
original spare parts are used and warranty claims
remain unaffected
.
● Pleasecomplywiththecountry-specificguidelines
regarding regular testing especially for the electrical
safety. ATMOS recommends a test every 24 months.
● Forrepair,thisdevicecanbereturnedtoATMOS.
● Beforereturningthedeviceforrepair,cleanandafter-
wards disinfect all secretion canister parts and hose
parts. The device´s surface also has to be disinfected.
● ATMOScannotguaranteeperfectfunctioningneither
will it be liable for damage to people or property if:
• Any non-original ATMOS parts are used,
• the user instructions given in this manual are not
followed exactly or are disregarded,
• assembly, resetting, alterations, extensions and
repairs are not carried out by people authorised by
ATMOS.
● Nowarrantyrightsshallexistintheeventofdamage
or failure caused by the use of non-ATMOS accessori-
es or non-ATMOS consumables.
● Inordertoprotecttheuser,theATMOSA/C161Bat-
tery must be in good condition with regard to technics
and hygiene prior to passing on. If there is a change of
patient or ownership then the device must be repro-
cessed acc. to Medical Devices Operator Ordinance,
Medical Devices Act and BVMed guidelines. An over-
sucked (contaminated) device must be repaired by
themanufacturer,byacertiedATMOSpartneroran
ATMOS authorized, specialized dealer.
● Payattentiontoregulationsandinstructionsvalidfor
the respective application range.
9.0 Maintenance
condensate
collector for a
quick view control
of a possible
contamination
9.2 Reprocessing
The way the suction device is used determines its
reliability and safety. These hygiene measures descri-
bed in the last chapter are indispensable for protecting
the patient and the user and for maintaining a safe and
reliable suction device.
When you can ensure the device was not
oversucked then perform a reprocessing
acc. to Medical Devices Operator Ordinance,
Medical Devices Act and BVMed guidelines.
The reprocessing consists of cleaning,
surface disinfection as well as the exchange of
consumables. An ATMOS set of consumables is
available for this purpose.
If you cannot exclude that the device was
oversucked then the device must be repaired by
themanufacturer,byacertiedATMOSpartner
or an ATMOS authorized, specialized dealer.
Subsequently the device may be operated again.
When is a suction unit oversucked?
A suction device is oversucked if suction material
penetrated into the interior of the device.
How can one realise that the suction device is
contaminated?
Generally a reduced suction capacity is a sign of possible
oversuction.
The ATMOS C 161 / C 261 Aspirator have a condensate
collectoratthebottomofthedevice(g
14
). For visual
inspection please remove the cap. The device is
oversucked if humidity or contamination is visible in the
condensate collector.
Measures
The device must be reprocessed by the manufacturer or
acertiedATMOSpartnerifthedeviceisoversuckedor
if any reservations exist regarding the hygiene condition.
A contaminated suction device bears a risk for the patient
as well as for the caregiver. Therefore, we recommend
regular checking of the condensate collector.
14
9.3 Sending in the device
1. Remove and properly dispose of consumables.
2. Clean and disinfect the products and accessories
according to the operating instructions.
3. Place used accessories with the device.
4. Fill in the form QD 434 „Delivery complaint / return
shipment“ and the respective Decontamination certi-
cate.
• This form is enclosed to each delivery and can be
found at www.atmosmed.com.
5. The device must be well padded and packed in suitable
packaging.
6. Place the form QD 434 „Delivery complaint / return
shipment“andtherespectivedecontaminationcerti-
cate in an envelope.
7. Axtheenvelopetotheoutsideofthepackage.
8. Send the product to ATMOS or to your dealer.