10.0 Trouble-shooting
Prior to dispatch, the ATMOS A / C 161 / 261 Aspirator was subjected to an extensive functional test. If, nevertheless, a failure
should appear, you may possibly clear it yourself if you follow these notes
:
Problem Possible causes Remedy
●Unitdoesnotstart Discharged battery Connect the battery charging
power pack to the device. The
battery should be recharged for
1 – 2 hours prior to operation
with battery.
Loose power plug of the charging device Check all plug and socket
connections. Pay attention to the
control lamp, it must be
illuminated when all connections
are o.k.
Insu󰀩cientperformance
Discharged battery Recharge the battery.
– Filter is blocked
– Exchangethelter.
● 1.Lowornovacuum
is indicated
–1.1DDSbacteriallterismissing –InsertDDSbacteriallter
– 1.2 Leakages within the hose system
or in collection canister lid
– Check collection canister lid and
hosesystemontighttting.
–Connectthelteronceagainto
the connection nozzle.
Check the suction lid on tight
andcorrecttting.
● 2.Highvacuumis
indicated
– 1.3 Secretion or blood has been
sucked in and valve plates of
the pump are contaminated
–2.1DDSbacteriallterisblocked.
Unit has to be returned for repair.
– ExchangeDDSbacteriallter
–2.2Floatofoverowsafetycloses
the collection canister inlet
– Check collection canister inlet; if
necessary empty secretion
canister, clean the oversuction
protectionandcheckoatball
forexibility.
When switching on,
the unit automatically
switcheso󰀨
Applied Vacuum (over 40%) Vent the vacuum, then restart