10.0 Trouble-shooting
Prior to dispatch, the ATMOS A / C 161 / 261 Aspirator was subjected to an extensive functional test. If, nevertheless, a failure
should appear, you may possibly clear it yourself if you follow these notes
:
Problem Possible causes Remedy
●Unitdoesnotstart – Discharged battery – Connect the battery charging
power pack to the device. The
battery should be recharged for
1 – 2 hours prior to operation
with battery.
– Loose power plug of the charging device – Check all plug and socket
connections. Pay attention to the
control lamp, it must be
illuminated when all connections
are o.k.
● Insucientperformance
– Discharged battery – Recharge the battery.
– Filter is blocked
– Exchangethelter.
● 1.Lowornovacuum
is indicated
–1.1DDSbacteriallterismissing –InsertDDSbacteriallter
– 1.2 Leakages within the hose system
or in collection canister lid
– Check collection canister lid and
hosesystemontighttting.
–Connectthelteronceagainto
the connection nozzle.
– Check the suction lid on tight
andcorrecttting.
● 2.Highvacuumis
indicated
– 1.3 Secretion or blood has been
sucked in and valve plates of
the pump are contaminated
–2.1DDSbacteriallterisblocked.
– Unit has to be returned for repair.
– ExchangeDDSbacteriallter
–2.2Floatofoverowsafetycloses
the collection canister inlet
– Check collection canister inlet; if
necessary empty secretion
canister, clean the oversuction
protectionandcheckoatball
forexibility.
• When switching on,
the unit automatically
switcheso
– Applied Vacuum (over 40%) – Vent the vacuum, then restart