9
5.0 Operation DDS
Important
notes on safety for the
DDS canister system
Do not operate the device without a
bacteriallter.Werecommendyou
always to store at least one spare
bacteriallter.
● Alwaysweardisposablegloveswhen
changingthebacteriallter!
● Exchangethebacteriallterwhenchang-
ing the patient. ATMOS recommends:
Replacethebacteriallterafter14days,
even if there is no patient change.
● Priortoeachusepleasecheckwheth-
erthebacteriallterisdryandclean.
Replacethebacteriallterwithanew
lterifitisdiscoloured,contaminated
oroversucked.Abacterialltermay
never be reused
● Vacuum connection
Direct-Docking-System
The vacuum connection between the pump
and the collection canister is created au-
tomatically as soon as the DDS canister is
positioned correctly.
5.1 DDS canister and
DDSbacteriallter
WiththeDDScollectioncanisteronarmsurface,position
the lid horizontally on top (the lid may not be twisted!)
Press down lightly onto the collection canister using both
hands until limit is reached (Fig 7).
5.2 Insertion/Removal of the
DDS canister
The lter must be xed to the housing.Afterwards the
secretion canister is inserted. For insert the DDS collec-
tioncanister,shiftithorizontallyontothebacteriallter;for
removal, pull it horizontally outside (Fig 8).
Please note that this order is strictly observed otherwise
power loss of the device could be the consequence!
5.3 Connect hose
Press the required DDS hose adapter with 6 or 10 mm
diameter into the hole of the DDS collection canister lid
twistingslightlytoensureatightt (Fig 10).
Twist slightly in the same manner when removing.
Fig 9.
Fig 8.
Fig 7.
Fig 10.
i
Tip
If required the canister can be ejected even easier from
the device by means of a lever instrument (e.g. plain
spattle); see Fig. 9.
!
!