14.0 Notes on EMC
14.3 Guidelines and Manufacturer´s Declaration - Immunity
TheATMOSA/C161/261Aspiratorisintendedforuseintheelectromagneticenvironmentspeciedbelow.The
customer or user of the ATMOS C 161 should ensure that it is used in such an environment.
Immunity Test
IEC 60601-
Test Level
Compliance Level
Electromagnetic Environment
- Guidance
Conducted RF
IEC 61000-4-6
3 V
e
150 kHz bis 80 MHz
10 V
Portable and mobile communications
equipment should be separated from the
RS 221 incl. the cables by no less than the
distances calculated/listed below.
Recommended distances:
d = 0,35 √P
d = 0,35 √P
80 MHz to 800 MHz
d = 0,70 √P
800 MHz to 2,5 GHz
where „P“ is the max. power in watts (W)
and D is the recommended separation
distance in meters (m).
Fieldstrengthsfromxedtransmitters,as
determined by an electromagnetic site (a)
survey, should be less than the compli-
ance level (b).
Interference may occur in the vicinity of
equipment containing following symbol
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz bis 2,5 GHz 10 V/m
NOTE 1 With 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2 These guidelines don´t like to be applicable in any case. The propagation of electromagnetic sizes is
inuencedbyabsorptionsandreectionsofbuildings,objectsandpeople.
aTheeldstrengthofstationarytransmitters,suchasbasestationsofcellularphonesandmobileterrainradio
equipment, amateur radio transmitters, cbm broadcast and TV stations cannot be predestined exactly.
To determine the electromagnetic environment in regard to stationary transmitters, a study of the location is to
beconsidered.IfthemeasuredeldstrengthatthelocationwheretheATMOSA/C161/261Aspiratoris
used exceeds the above compliance level, the ATMOS A / C 161 / 261 Aspirator is to be observed to verify
the intended use.
If abnormal performance characteristics are noted, additional measures might be necessary, e. g. a changed
arrangement or another location for the device.
bWithinthefrequencyrangeof150kHzto80MHztheeldstrengthistobebelow3V/m.