Reprocessing
The product may become damaged.
Use only lint-free, soft cloths.
Use only brushes with soft bristles.
Never clean and/or disinfect the product in an ultrasonic bath.
Follow the corresponding operating instructions of all aids, tools, and devices
used.
Unsuitable cleaning agents and disinfectants
The product may become damaged.
Do not use any process chemicals containing the following ingredients on the ATMOS
Scope as well as for all accessories:
Chloramines
Phenol derivatives
Incorrect cleaning and disinfection
Corrosion due to moisture.
Remove the products immediately after the program is nished.
Never clean and/or disinfect the product in an ultrasonic bath.
Avoid excessive pressure and tension on the exible endoscope components.
Only place the device in suitable load carriers with silicone pads in order to avoid
damage to the device. Please clean carefully, as the tube and bendable end
can be damaged by kinking, twisting, pushing, or pulling. Gross contamination
should be removed carefully with a soft cloth or a soft brush.