42
Troubleshooting
7 Troubleshooting
Error symptom Possible cause Remedy
Cloudy image Glass surfaces are
dirty
Clean glass surfaces in accord-
ance with chapter “5.3 Preparing
surfaces” on page 31
Leaky, defective
lens system
Send in the ATMOS Scope for
repair
Image is too dark,
insucient illumina-
tion
Glass surfaces are
dirty
Remove residues according to
chapter “5.3 Preparing surfaces”
on page 31; check water
quality
Stubborn residue on
the glass surfaces
Clean again, rub thoroughly if
necessary
LED is defective Send in the ATMOS Scope for
repair
Brightness is set
incorrectly
Adjust brightness in the software
settings