Reprocessing
Rapicide PA, Cantel Hydrogen peroxide 22 %
Acetic acid 9 %
Peroxyacetic acid 5 %
Liquid concentrate
5.3.3 Manual cleaning and disinfection
Cleaning:
Temp.: 30–35 °C, 0.8 % Cidezyme, drinking
water
Soft cloth
Soft nylon brush 1 min
(Brush length: 35 mm
Brush width: 7 mm
Total length: 175 mm
Bristles: nylon
Hardness: soft)
Disposable syringe (20 ml), rinse at least 5
times
Completely immerse in cleaning solution.
Ensure that all accessible surfaces are moistened.
With the insertion tube lying in the solution, clean it with a soft
cloth and clean the ATMOS Scope handle with a soft cleaning
brush until no residue is visible on the surface.
Move the movable component (deection lever) 5 times in each
direction as far as it goes during cleaning.
Using a disposable syringe, thoroughly rinse these places with the
cleaning solution at least 5 times.
Do not use metal cleaning brushes or other abrasives that could
damage the product surfaces, as they could cause corrosion.