36
Reprocessing
Intermediate rinsing:
• Drinking water
• Water jet gun for 60 s
• Fill the sink with enough drinking water so that the ATMOS Scope
is completely immersed.
• Set the pressure for the water jet gun to maximum 2 bar.
• Using the water jet gun, rinse o the entire handle for 60 seconds
and pay particular attention to gaps and hidden surfaces.
Recommendation:
Rinsing should be done under water to avoid splashing
contaminated liquid
.
Final rinse:
• 1min
• Cold drinking water
• Move 5 times
• Rinse the probes for 1 minute under cold running drinking water
to remove any cleaning agent residues.
• Move the deection lever at least 5 times in each direction as far
as possible.
Disinfection
Disinfection:
• Temp.: at least 20–25 °C, Cidex OPA, drinking
water
• 12min
Completely immerse the ATMOS Scope in the disinfection solution.
Final rinse:
• Temp.: 30–45 °C, Time 3 min
• Water quality: drinking water
Drying:
• Temp.: 30–45 °C
• 3min
Please observe the shelf life of the immersion disinfection, according to the manufacturer’s instructions.