Reprocessing
5.3.4 Mechanical cleaning and disinfection
General notes for mechanical cleaning
Inspect visible surfaces for residual contamination after mechanical cleaning/disin-
fection. Repeat the cleaning process if necessary.
The product is compatible with several washer-disinfectors for exible endoscopes.
For further details regarding operation, please refer to the operating instructions for
the washer-disinfector.
If it is unclear whether the product can be cleaned and disinfected with the existing
washer-disinfector for endoscopes, please contact the manufacturer of the wash-
er-disinfector for endoscopes and, if necessary, check with the manufacturer which
cleaning and disinfection program is suitable for the product.
Product damage due to excessive temperatures. During mechanical cleaning and
disinfection, the temperature should not exceed 65 °C / 149 °F.
Position and gently x the endoscope in a cleaning basket lined with silicone
mats in the washer-disinfector in accordance with the machine manufacturer’s
instructions.
Connect the tube for the leakage tester to the pressure compensation valve 2 of
the ATMOS Scope handle and use an adapter if necessary.
The validation that the product can be cleaned and disinfected has been conducted
using the washer-disinfector for exible endoscopes from Belimed Type WD 430 with
the rack Endoscopy 2 level.