38
Reprocessing
Pre-cleaning:
Drinking water, wipe 5 times
Soft nylon brush, 1 min
(Brush length: 35 mm
Brush width: 7 mm
Total length: 175 mm
Bristles: nylon
Hardness: soft)
Step 1:
Remove any gross contamination with a cloth dampened in
cold drinking water. Wipe the insertion tube 5 times from
the handle toward the tip.
Step 2:
Brush the handle using a nylon brush with soft bristles
for 1 minute under running drinking water. Pay particular
attention to crevices and hidden areas.
Mechanical cleaning and disinfection:
Pre-rinse: 3min
Drain
Cleaning: 7 min, 55 °C / 131 °F with 5 ml
neodisher® Mediclean forte
Drain
Rinse: 2 min, 55 °C with drinking water
Drain
Accessories for storage: screen basket,
silicone mat
Lay the ATMOS Scope on the silicone bubble mat and x
the insertion tube at a minimum of 2 places.
Select a suitable program for cleaning and disinfection
with the adjacent parameters.
Washer-disinfector: in accordance with EN ISO 15883-1.
Comments: The specied temperatures and times refer to the water
supply and the set value of the washer-disinfector.
Temperature of cold drinking water: 10–25 °C and of cold
deionized water: 10–25 °C.