USER MANUAL 39
Troubleshooting:
Display indication
Possible cause
Remedy
Initial heating Td
temperature exceeds
105°C, the sensor
does not react when
touched with a finger
The temperature sensor is
damaged.
Contact an authorised service
point.
Initial heating Td
temperature does not
reach 105°C during 3
minutes time,
the halogen heater(s)
do not light.
The heater is damaged.
Replace the heater.
"Test ..."
Auto-tests in progress /
electronic unit damage
wait for 1 minute
" - - - - "
The moisture analyzer is
during zeroing
/ mechanical damage
wait for 1 minute
check if the moisture analyzer
is placed on stable ground, not
affected by vibrations
„Tare range
exceeded”
Tare key pressed during zero
indication
Moisture analyzer indications
must be different than zero
„Zeroing range
exceeded”
Permissible zeroing range
was exceeded
Remove the load from the pan
Weighing range
exceeded”
Permissible weighing range
(Max +9e) was exceeded
Reduce the load
„Measuring range
exceeded (+)”
Upper limit of analog-digital
transducer measuring range
was exceeded
Remove the load from the pan
„Measuring range
exceeded (-)”
Lower limit of analog-digital
transducer measuring range
was exceeded
Check if there are all necessary
pan elements