USER MANUAL 39
Troubleshooting:
Initial heating Td
temperature exceeds
105°C, the sensor
does not react when
touched with a finger
The temperature sensor is
damaged.
Contact an authorised service
point.
Initial heating Td
temperature does not
reach 105°C during 3
minutes time,
the halogen heater(s)
do not light.
Auto-tests in progress /
electronic unit damage
The moisture analyzer is
during zeroing
/ mechanical damage
wait for 1 minute
check if the moisture analyzer
is placed on stable ground, not
affected by vibrations
Tare key pressed during zero
indication
Moisture analyzer indications
must be different than zero
Permissible zeroing range
was exceeded
Remove the load from the pan
„Weighing range
exceeded”
Permissible weighing range
(Max +9e) was exceeded
„Measuring range
exceeded (+)”
Upper limit of analog-digital
transducer measuring range
was exceeded
Remove the load from the pan
„Measuring range
exceeded (-)”
Lower limit of analog-digital
transducer measuring range
was exceeded
Check if there are all necessary
pan elements