AUDIO OUTPUT: This output can be used in conjunction with AUDIO INPUT, as
described above. It can also drive headphones or a remote powered speaker, like
the front panel Headphones Socket.
SYNC OUTPUT: The signal available at this BNC connector is intended to be used
as an External trigger for an oscilloscope when Seal Test (in V-Clamp), Tuning
(in I-Clamp) or Pulse is activated. The Sync Output is a 0 to 3.4V step aligned
with the onset of the Seal Test, Tuning or Pulse step.
SIGNAL GROUND: This 4 mm socket is the primary grounding point for the
MultiClamp 700A. Ideally, it should be connected to a grounding bus to which is
also connected the Faraday cage and the signal grounds of other instruments used
in your experiment.
DEVICE ADDRESS: This is used to specify the device number of the MultiClamp
700A. The number is entered in the Select Device dialog box when first starting
up the MultiClamp Commander (or by using the Select Device toolbar button).
SCREW CONNECTOR (CHASSIS GROUND): This provides an alternative chassis
or power supply ground.