EasyManua.ls Logo

Axon MultiClamp 700A - Zap

Default Icon
159 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
132 Reference Section
MultiClamp 700A Theory and Operation, Copyright 2000, 2001 Axon Instruments, Inc.
Zap
Zap applies a large, brief voltage pulse to the electrode when in V-Clamp mode, to
facilitate breaking into a cell for whole-cell recording.
Zap is triggered by pressing the
button in the V-Clamp pane.
The conventional method for rupturing a membrane patch to go to whole-cell
recording is to apply a pulse of suction. Sometimes this method damages the cell. Zap
provides an alternative method. It applies a large (1 V) voltage pulse that ruptures the
patch, presumably by causing dielectric breakdown of the membrane. The Zap
duration can be varied; it is best to use the minimum duration that is likely to achieve
the desired result, because too long a Zap could cause the seal resistance to deteriorate.
A duration of 0.5 or 1 ms is suggested for initial attempts.
Apply a repetitive test pulse (e.g. Seal Test) and press the Zap button while carefully
monitoring Membrane Current. Sometimes it helps to apply steady suction while
Zapping. Successful break-through is signaled by an increase in the current noise and
by large whole-cell capacitance transients.

Table of Contents

Related product manuals