Oscilloscope Triggering
SYNC output on the rear panel of the MultiClamp 700A provides a signal for
triggering an oscilloscope (or for triggering in Clampex).
See also Input/Output Connections.
The signal available at this BNC connector is intended to be used as an External trigger
for an oscilloscope when Seal Test (in V-Clamp), Tuning (in I-Clamp) or Pulse is
activated. The Sync Output is a 0 to 3.4V step aligned with the onset of the Seal Test,
Tuning or Pulse step.