Specifications 139
Chapter 7
Output Gain
Post-filter gain of 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, 2000.
Audio Monitor
The Audio Monitor output can select Current, Voltage or Voltage x 100 for either
Channel 1 or Channel 2. The selected signal is available for direct monitoring or via a
voltage-to-frequency converter (VCO). The VCO ranges from ~4000 Hz @ +100 mV
to ~300 Hz at –100 mV.