Index g
(Master – to be added to the documentation)
M688 70 01 05 F04 / 3891 7319 (GB) Sheet 2 of 3
1-Technical Safety Check (TSC)
EN5- 2 Perfusor® Space, 3.0
Operation
Syringe fastening
Syringe recognition
Infusion
Buttons on the operating unit
Staff call
Trigger bolus at the device
Trigger bolus by pressing the PCA
button
PCA lock (bottom side of pump) with
inserted 50/60 ml syringe
PCA lock Syringe Anti Removal CAP PSP
with inserted 50/60 ml syringe
Switch on unit without power supply
Self-test
Magnetic function of the battery
compartment cover
Pressure cut-off
Syringe type: “#LehreOPS50”
Delivery rate: 200 ml/h
- Strain gauge pressure measurement
Pressure stage 1 (9…15N)________N
Pressure stage 3 (25…33N)_______N
Pressure stage 8 (63…76N)_______N
Error message “Alarm / pressure too
high” at every pressure stage
- Motor power limitation
Pressure stage 1 (11…28N)________N
Pressure stage 3 (30…49N)________N
Pressure stage 6 (58…80N)________N
Error message “Alarm / drive blocked”
at every pressure stage
Visual Inspection Electrical Safety
According to IEC / EN 60601-1
or VDE 0750 and VDE 0751
Functional Inspection