Index g
(Master – to be added to the documentation)
M688 70 01 05 F04 / 3891 7319 (GB) Sheet 2 of 3
1-Technical Safety Check (TSC)
EN5- 2 Perfusor® Space, 3.0
Operation
❒ Syringe fastening
❒ Syringe recognition
❒ Infusion
❒ Buttons on the operating unit
❒ Staff call
❒ Trigger bolus at the device
❒ Trigger bolus by pressing the PCA
button
❒ PCA lock (bottom side of pump) with
inserted 50/60 ml syringe
❒ PCA lock Syringe Anti Removal CAP PSP
with inserted 50/60 ml syringe
Switch on unit without power supply
❒ Self-test
❒ Magnetic function of the battery
compartment cover
Pressure cut-off
Syringe type: “#LehreOPS50”
Delivery rate: 200 ml/h
- Strain gauge pressure measurement
❒ Pressure stage 1 (9…15N)________N
❒ Pressure stage 3 (25…33N)_______N
❒ Pressure stage 8 (63…76N)_______N
❒ Error message “Alarm / pressure too
high” at every pressure stage
- Motor power limitation
❒ Pressure stage 1 (11…28N)________N
❒ Pressure stage 3 (30…49N)________N
❒ Pressure stage 6 (58…80N)________N
❒ Error message “Alarm / drive blocked”
at every pressure stage
Visual Inspection Electrical Safety
According to IEC / EN 60601-1
or VDE 0750 and VDE 0751
Functional Inspection