Index g
(Master – to be added to the documentation)
M688 70 01 05 F04 / 3891 7319 (GB) Sheet 1 of 3
Perfusor® Space, 3.0 5- 1EN
5-Technical Safety Check (TSC)
Perfusor® Space
Checklist for Technical Safety Check – Every 24 Months
Unit: Perfusor®Space
Manufacturer: B. Braun Melsungen AG
User
Observe the Service Manual and the instructions for use. All measured values are to be documented.
Accessories used should be included in testing. Make exclusive use of calibrated measuring equip-
ment.
Article No. Unit No. Year of Procurement Stock No.
Visual Inspection Electrical Safety
According to IEC / EN 60601-1
or VDE 0750 and VDE 0751
Functional Inspection
Perfusor® Space:
Cleanliness, completeness, damage and
defects affecting safety, damage and
readability of the labels.
Particularly:
Syringe holder with blade
Syringe fastening
Membrane in drive head
Axial clearance of drive
Screw covers
Connectors “P2” and “P3”
Protection cap at drive head
Accessories:
Cleanliness, completeness, damage and
faults affecting safety, damage and
readability of the label
Check the unit and the accessories for
compatibility
The patient and housing leakage current of
the Perfusor® Space is caused exclusively by
the operating voltage supply (Power Supply
SP or SpaceStation).
The Technical Safety Checks of the power
supply SP (drawing No. M001321005 F04)
or of the SpaceStation (drawing No.
M690000046 F04) serve to check whether
both limit values are met.
Locking with second unit
Operating unit magnets
Battery compartment cover
Determine age of battery
If older than 4 years:
recommend replacement of battery
or perform battery check
(refer to “Procedural Instructions
on the TSC“)
Switch on unit with power supply
Self-test
Indicator lamps (LEDs)
Audible alarm
Visual alarm
Status display
Lighting
Battery alarm
5 - 4