EasyManua.ls Logo

B. Braun Perfusor Space - Battery Alarm

B. Braun Perfusor Space
10 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Procedural Instructions on the TSC
6
Perfusor® Space, 3.0 6- 5EN
7. Trigger bolus by pressing the PCA button.
For this purpose, select a PCA drug from the database, fit the
PCA button to P2, and trigger bolus.
Confirm that the green LED on the PCA button lights up. Bo-
lus is delivered.
8. Staff call
a) Plug service connector SP on connector “P2”.
b) Open syringe holder while the infusion is administered.
The red LED on the service connector SP lights up.
9. PCA lock on the Perfusor® Space
a) Check the PCA lock with an approved 50/60ml syringe.
10. PCA lock of the Syringe Anti Removal Kit PSP
a) Place on the drive head and lock.
b) Check the Syringe Anti Removal Kit PSP with an ap-
proved 50/60ml syringe.
c) Unlock and remove from the drive head.
Battery Alarm
1. Switch device off.
2. Pull off the power supply.
3. Switch on unit.
Self-test is carried out.
4. Open battery compartment cover during operation.
An alarm is activated.
5. Remove battery.
An alarm sounds with the piezo buzzer for at least 3 minutes.

Other manuals for B. Braun Perfusor Space

Related product manuals