Procedural Instructions on the TSC
6
Perfusor® Space, 3.0 6- 3EN
Functional Inspection Perfusor® Space
Mechanical Inspection
1. Fit the unit to be tested on top of another Space device and
check the proper functioning of the lock.
2. Fit the unit to be tested under another Space device and
check the proper functioning of the lock.
3. Hold the device with the operating unit downwards. The op-
erating unit must not open.
4. Check the battery compartment cover lock for proper opera-
tion.
Functional Check
1. Determine age of battery
Open battery compartment cover and remove battery.
Read the batch number printed on the type plate of the
battery and determine the age of the battery.
a) For NiMH batteries, see Fig.: 6 - 2
b) For Li-Ion batteries, see Fig.: 6 - 3
Note
If the battery is older than 4 years recommend the replacement of
the battery or perform the battery check.
Fig.: 6 - 2 NiMH batteries Fig.: 6 - 3 Li-Ion batteries
Letter: year of production
according to left table
Two digits: calender week
Example:
Produced in week 42/2008
D 2003
E 2004
F 2005
G 2006
H 2007
I 2008
...
N2013
First two digits: calender week
Second two digits: year of production
Example:
Produced in week 41/2011