Conductivity I 34
Where:
K = Cell constant
C
std
= Value of conductivity standard solution
C
meas
= Measured value
G = Raw cell constant (0.1, 1 or 10)
Calculating the Temperature Coefficient
1. Do not connect the temperature probe to the meter.
2. Press and hold the °C key to enter the temperature setting.
3. Press the / key to set the temperature to 25° C and press
the Enter key to confirm.
4. Place the conductivity electrode into the sample solution, record
the temperature value T
A
and conductivity value C
TA
.
5. Condition the sample solution and electrode to a temperature T
B
that is about 5 to 10° C different from T
A
. Record the conductivity
value C
TB
.
6. Calculate the temperature coefficient using the formula below.
T
C
= [C
TB
C
TA
] / [C
TA
(T
B
25) C
TB
(T
A
25)]
Where:
T
C
= Temperature coefficient
C
TA
= Conductivity at temperature A
C
TB
= Conductivity at temperature B
T
A
= Temperature A
T
B
= Temperature B
Calculating the TDS Conversion Factor
To determine the TDS factor of sample solution use the formula below.
Factor = Actual TDS / Actual Conductivity @25° C
Where:
Actual TDS = value from the high purity water and precisely weighed
NaCl or KCL reagent
Actual Conductivity = the meter measured conductivity value
For example:
Dissolve 64 grams of the potassium chloride (KCl) reagent in 1 liter
distilled water. If measured conductivity is 100 mS/cm, then TDS factor
is 0.64.
Conductivity to TDS Conversion Factors
Conductivity at 25° C
TDS (KCl)
TDS (NaCl)
ppm
Factor
ppm
Factor
84 µ S/cm
40.38
0.5048
38.04
0.4755
1413 µ S/cm
744.7
0.527
702.1
0.4969
12.88 mS/cm
7447
0.5782
7230
0.5613
Optional Accessories
Conductivity Electrodes
Order Code
Description
CON-0.1
For measuring the pure water
CON-1
For general purpose applications
CON-10
For measuring the high conductivity liquids
Temperature Probe
Order Code
Description
TP-10K
Range: 0 to 100° C (32 to 221° F), 1 m (3.3 ft.) cable
Solutions
Order Code
Description
ECCS-84
Conductivity standard solution 84 µ S/cm, 480 ml
ECCS-1413
Conductivity standard solution 1413 µ S/cm, 480 ml
ECCS-1288
Conductivity standard solution 12.88 mS/cm, 480 ml
ECCS-5000
Conductivity standard solution 50.00 mS/cm, 480 ml
ECCS-1118
Conductivity standard solution 111.8 mS/cm, 480 ml
Communication and Power Supply
Order Code
Description
USB-B
USB connector A to B, 1 m (3.3 ft.) cable
DCPA-5V
DC 5V power adapter, european standard plug

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bante Instruments 9 Series and is the answer not in the manual?

Bante Instruments 9 Series Specifications

General IconGeneral
BrandBante Instruments
Model9 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals