EasyManuals Logo
Home>Bante Instruments>Measuring Instruments>9 Series

Bante Instruments 9 Series User Manual

Bante Instruments 9 Series
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #19 background imageLoading...
Page #19 background image
ORP I 18
Appendix
Preparation of ORP Standard Solutions
Quinhydrone solution A: Dissolve 3 grams of quinhydrone reagent in
500 ml of the pH 4.01 buffer solution, stir the solution for 10 minutes.
Undissolved quinhydrone reagent must be present. If necessary, add
the reagent.
Temperature
Potential (± 10 mV)
20° C
268 mV
25°C
263 mV
30° C
260 mV
Quinhydrone solution B: Dissolve 3 grams of quinhydrone reagent in
500 ml of the pH 7.00 buffer solution, stir the solution for 10 minutes.
Undissolved quinhydrone reagent must be present. If necessary, add
the reagent.
Temperature
Potential (± 10 mV)
20° C
94 mV
25°C
87 mV
30° C
80 mV
Due to the quinhydrone solution is susceptible to air oxidation in
storage, make sure to prepare the fresh solution before use.
Preparation of Electrode Storage Solution
Dissolve 29.8 grams of analytical grade potassium chloride (KCl)
reagent in 100 ml distilled water.
Add pH 4.01 standard buffer and adjust solution to pH 4.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bante Instruments 9 Series and is the answer not in the manual?

Bante Instruments 9 Series Specifications

General IconGeneral
BrandBante Instruments
Model9 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals