Dissolved Oxygen I 38
Dissolved Oxygen Measurement
The 9 series meter can be used to measure the water, wastewater,
brine and other liquids. If your sample is seawater or water containing
large amounts of salt, make sure to set the salinity coefficient before
measurement. Some gas and steam such as chloride, sulfur dioxide,
sulfureted hydrogen and carbon dioxide can permeate the membrane
via diffusion. Their existence will influence the measurements. If the
sample contains the solvent, grease, sulfide and alga, the membrane
will be damaged or eroded.
1. Set the barometric pressure and salinity coefficient in the setup
menu.
2. Rinse the dissolved oxygen electrode with distilled water. Place
the electrode into the sample solution and stir gently.
3. If the Auto-Hold option in the setup menu is enabled, the meter
will automatically sense a stable reading and lock measurement,
the icon appears on the display. Press the key to resume
measuring.
If the option is disabled, the meter will continuously measure and
update the readings.
4. Wait for the measurement to stabilize and record the reading.
If the meter shows ---- indicating the measurement exceeds the
range, remove the electrode from the sample immediately.
To record the measurement at the predefined time intervals, refer
to the Interval Readings section on page 40.
Electrode Maintenance
Rinse the dissolved oxygen electrode thoroughly with distilled
water after use.
Do not touch the membrane and always keep it is clean and wet.
If you do not use the electrode for long periods, screw off the
membrane cap and rinse the electrode anode, cathode, membrane
cap with distilled water and blot dry. Install the electrode and
store dry.
Appendix
Preparation of Zero Oxygen Solution
Dissolve 500 mg of the sodium sulfate (Na
2
SO
3
) reagent and a small
amount of cobalt (II) chloride hexahydrate (CoCl
2
6H
2
O) in the 250 ml
distilled water, mix the solution until reagent is completely dissolved.
Preparation of Air-Saturated Water
Use an air-pump to blow air into distilled water at least 1 hour, while
stirring the solution.
Optional Accessories
Dissolved Oxygen Electrode and Components
Order Code
Description
DO100
Dissolved oxygen electrode, range: 0 to 20 mg/L
DO-MEM
Membrane cap, 2 PCS/set
Solution
Order Code
Description
DO-ES
Electrolyte solution, 30 ml
Communication and Power Supply
Order Code
Description
USB-B
USB connector A to B, 1 m (3.3 ft.) cable
DCPA-5V
DC 5V power adapter, european standard plug
Cathode
Anode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bante Instruments 9 Series and is the answer not in the manual?

Bante Instruments 9 Series Specifications

General IconGeneral
BrandBante Instruments
Model9 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals