SINGLE NEEDLE | Start the Treatment
Once the treatment has been started, press the
flashing arterial pressure control
to centralize
the alarm limits around the current value.
The alarm limits for the venous pressure are
automatically set
once the treatment has started
to +350mmHg (high limit) and +150mmHg (low limit). If it is necessary to change the limits,
press the venous pressure control. Alternately the limits can also be changed by in the blood
menu low and/or high venous pressure limit.
The stroke volume that can be achieved is determined by the size and position of the venous
pressure alarm window, and the potential expansion volume provided by the expansion
chamber.
HC10139 |