HYGIENIC MAINTENANCE | Changing the Ultrafilter
1. Pull the latch and press downwards
2. Pull the ultrafilter gently downwards and remove it
3. Label the new ultrafilter with the actual date
4. Lubricate the ultrafilter connections with RO water
5. Insert the new ultrafilter into the holder and push it gently upwards
6. Push the lower latch into position and close the holder until a ‘click’ is heard
7. Press the
disinfection button
, under the tools tab and
confirm the UFD filter replacement
.
8. After changing the ultrafilter, a
heat disinfection
must be performed before a new
treatment.