TROUBLESHOOTING | Power Failure
If there is a power failure, the
machine has a
battery back-up which lasts approximately 30
minutes
and keeps the screen and the blood
pump operational. During this time the dialysis
fluid is not heated and the dialysis fluid is in
bypass.
If all the battery back-up charge is used up or, for
some reason it does not work, then the machine
will shut down. All the settings and the actual
values will be kept.
When the power returns,
press the On/Off button
to start the machine. The machine will perform a
recovery and the treatment will continue from
where it stopped.
However,
all the treatment
parameters must be checked when starting after
a recovery.
Battery connect indicator is lit green
if the
battery back-up has been installed
Battery charge indicator is lit yellow
when
the mains cable is connected to the mains
supply and the main switch is switched on