TROUBLESHOOTING | Power Failure
• If there is a power failure, the
machine has a
battery back-up which lasts approximately 30
minutes
and keeps the screen and the blood
pump operational. During this time the dialysis
fluid is not heated and the dialysis fluid is in
bypass.
• If all the battery back-up charge is used up or, for
some reason it does not work, then the machine
will shut down. All the settings and the actual
values will be kept.
• When the power returns,
press the On/Off button
to start the machine. The machine will perform a
recovery and the treatment will continue from
where it stopped.
However,
all the treatment
parameters must be checked when starting after
a recovery.
Battery connect indicator is lit green
if the
battery back-up has been installed
Battery charge indicator is lit yellow
when
the mains cable is connected to the mains
supply and the main switch is switched on
HC10139 |