114
HYGIENIC MAINTENANCE | External Components
After each dialysis treatment
, the machine exterior
surfaces must be disinfected.
Using a cloth moistened with either ethanol 70%,
isopropanol 60% or 1% hypochlorite,
wipe all of the
machines exterior surfaces and the top tray.
In case of cleaning with
hypochlorite
, wipe afterwards
with water.
The
pick-up tubes
need to be flushed inside and out with
water between each treatment, and then returned to the
holder to dry naturally. Once a week, repeat exactly the
same procedure using ethanol 70% instead of water.