PREPARING THE MACHINE | Manual Priming
In
manual priming
the operator is in charge of the priming process.
Priming of the blood side of the circuit can start as soon as the blood pump button is flashing, and
the blood lines have been attached.
In this case, the caps from the dialyzer blood ports should be moved to the dialyzer fluid ports to
prevent spillage.
Alternatively, priming can also be delayed so that it is completed once the machine has reached
full conductivity, as displayed by the green flow path.
In this situation the dialysis fluid tubes of the machine can be connected to the dialyzer, while
remaining in bypass, allowing the blood side priming to be started and preventing spillage.
The dialyzer is then rotated and the machine taken out of bypass at the appropriate point.