44
PREPARING THE MACHINE | Extra Priming & Recirculation
• It is possible after the completion of either manual or assisted priming, to
perform extra priming
or to
recirculate the priming solution
.
• For both procedures the operator must ensure that there is enough priming solution to complete the action.
• For recirculation, both ends of the blood lines must be connected together using a sterile accessory.
• The priming fluid should be connected to the circuit via an infusion line to compensate for the ultrafiltration
that occurs in the dialyzer during the recirculation phase.
• When recirculation is finished or the blood pump is stopped, close the clamp on the infusion line, then press
Connect patient
HC10139 |