HEMODIALYSIS TREATMENT | Pause the Treatment
1. Press the
functions button
to open the functions menu
2. Press
pause treatment
3. Select pause with or without blood return and press OK
4. Follow the instructions on the screen to proceed
5. To resume the treatment, press the blood pump button to stop
the blood pump
6. Select connect patient, connect the blood lines to the patient
and press confirm
7. Press the blood pump button and adjust the blood flow
8. Press the ultrafiltration button to resume UF removal
When the treatment is paused:
The blood flow rate is decreased and the arterial and venous pressure alarm windows are widened
The elapsed time in pause is displayed and a sound alarm is triggered every 15 minutes
Diascan
measurements, Isolated UF and profiling is disabled
Continuous heparinization administration and automatic BPM is deactivated but heparin bolus dose and
manual BPM is enabled