FEATURE | Profiling Ultrafiltration
With linear profiles the machine knows the set time and UF
volume, the operator sets the
start UF rate
, and the machine will
calculate a smooth decreasing profile.
With step profiles the machine knows the set time and UF volume,
the operator sets the
start UF rate
, and the
number
of steps
and the machine will calculate the profile.
With interval profiles the machine knows the set time and UF
volume, the operator sets the
start UF rate
, and the
number of
intervals
and the machine will calculate the profile.