EasyManua.ls Logo

baxter HomeChoice PRO - Page 168

baxter HomeChoice PRO
377 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 10-11
10. Operating Instructions – Make Adjustments
* The settings for these percentages are calculated for you in Table 19-8 on
page 19-17 in 19.16, Determining Initial Drain Alarm Volume Settings.
Table 10-1. Recommended Starting Point for I-Drain Alarm Setting
Last Fill Solution Last Fill Dwell Time % of Last Fill Volume
Dianeal 8 to 16 hours 70%*
Dianeal 2 to 4 hours 85%*
Extraneal 8 to 16 hours 95%*
h
NOTE:
If the I-DRAIN ALARM is set to OFF, the HomeChoice APD System
moves on to Fill when a No Flow condition occurs. Baxter
recommends that this setting not be used since there is no Minimum
Drain Volume requirement. A LOW DRAIN VOLUME alarm is posted if
the flow rate is below 50 mL/min for 10 minutes. Subsequent alarms
are posted every 5 minutes. The I-DRAIN: OFF option is not available
in the Low Fill Mode.
h
NOTE:
For “Dry Day” patients, set the I-Drain Alarm to 0 mL, or a very small
volume. If your patient line is greater than 12 feet (3.6 meters) in
length, set the I-Drain Alarm to at least 30 mL. Check with your
dialysis center for the correct number to set. Setting the I-Drain Alarm
to NO in the Standard Fill Mode will activate a No Flow or Slow Flow
alarm.

Table of Contents

Related product manuals