EasyManua.ls Logo

baxter HomeChoice PRO - Perform a Hi-Dose Day Exchange

baxter HomeChoice PRO
377 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12. Operating Instructions – Perform Therapy
12-16 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
12.7 Perform a Hi-Dose Day Exchange
Follow the steps below to perform daytime exchange.
Average Dwell
Time
Average actual Dwell Time only reflects the average of the
nighttime cycles. During daytime cycles, 0:00 appears.
Press
ENTER to review cycle-by-cycle information. Press
STOP to return to the previous menu.
Review Program
Allows you to review the programmed treatment.
Current Time
Current time.
h
NOTE:
If you bypass a Hi-Dose daytime exchange, the solution volume from
that exchange will be added to the available nighttime therapy
volume.
Steps to perform a daytime exchange Display screen
Hi-Dose therapy begins when INITIAL DRAIN is
complete.
1. DAY FILL begins automatically when INITIAL
DRAIN is complete.
Table 12-6. Options Available During Day Dwell (Continued)
Option Description
AVG DWELL TIME: O:OO
REVIEW PROGRAM
6:10 PM
INITIAL DRAIN
DAY FILL 1 OF 1

Table of Contents