07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 18-29
18. Troubleshooting
18.3.14 Caution: Positive UF
CAUTION: POSITIVE UF
Message Display:
Low Fill Mode only.
Cause: You have removed more than the allowable ultrafiltrate
volume in either the current cycle or over the course of
several cycles.
The allowable volume can be set between 0 (zero) and
5000 mL.
This is a Manual Restart Alarm.
To Correct: 1. Press
STOP to mute the alarm.
2. Press
V until BYPASS appears.
3. Press
ENTER to choose BYPASS.
You have bypassed the alarm and treatment resumes.
Notify your dialysis center if the alarm continues.
Ultrafiltration (UF) may be set too high, or the alarm setting
may be too low.