18. Troubleshooting
18-30 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
18.3.15 Verify I-Drain
VERIFY IDRAIN
Message Display:
Cause 1: The I-DRAIN ALARM setting is OFF.
To Correct
Cause 1:
1. Press the power switch OFF and ON again.
2. Set the I-DRAIN ALARM to a value other than OFF. See
10.2.7, I-Drain Alarm, on page 10-9.
h NOTE: You have permanently changed the Initial Drain Alarm setting for
this therapy and all future therapy sessions.
Cause 2: The I-Drain Alarm setting is lower than expected.
To Correct
Cause 2:
1. Press
STOP and the setting blinks.
2. Press
U or V to temporarily change the minimum
I-Drain Alarm setting.
3. Press
ENTER to accept the value.
4. Press
GO to resume Initial Drain.
h NOTE: You have temporarily changed the Initial Drain Alarm setting for this
treatment session. To permanently change the I-Drain Alarm, see
10.2.7, I-Drain Alarm, on page 10-9.
VERIFY IDRAIN: OFF