4-1
DMR V4C 3.3 (User Manual) V-4 06-May-2022
4 Installation and Assembly
WARNING:
Only use the V4C-O2E Oxygen Mixer with compatible ventilators listed in Section 2.4,
Compatible Ventilators and marked with an "O2E" port. Do not attempt to use the Oxygen Mixer
with other ventilators.
WARNING:
Do not use unauthorized accessories with the V4C-O2E Oxygen Mixer system. Use of incorrect
accessories can lead to malfunction or damage to the device.
WARNING:
Do not place anything in the space between the bottom of the ventilator and the top of the V4C-
O2E Oxygen Mixer. Cooling vents located in this space may become compromised.
WARNING:
Do not use the V4C-O2E Oxygen Mixer if it is not appropriately affixed to the ventilator. See
Section 4.1, Installation and Assembly for instructions on how to securely attach the V4C-O2E
Oxygen Mixer to the ventilator.
WARNING:
If any of the assembly steps in Section 4.1, Installation and Assembly cannot be completed do
not use the V4C-O2E Oxygen Mixer. Contact your customer service representative.
WARNING:
Never use the V4C-O2E Oxygen Mixer or components that appear to be damaged or faulty.
Contact Baylis Medical Technologies if the cause of a problem cannot be determined.
WARNING:
Do not connect high pressure oxygen directly to the ventilator. Always follow the instructions on
the compatible ventilator when using its low flow oxygen port to deliver oxygen.