Electromagnetic Emissions and Immunity
12-2 User Manual
DMR V4C 3.3 (User Manual) V-4 06-May-2022
Table12-3.Electromagnetic immunity — conducted and radiated RF
Phenomenon
Basic EMC Standard /
Test Method
Immunity Test Levels for Home
Healthcare Environment
Radiated Field Immunity
IEC 60601-1-2 / IEC 61000-4-3
10 V/m (80% AM at 1KHz sine)
80 MHz – 2.7 GHz
Proximity Fields from RF Wireless Com-
munications Equipment
Proximity Fields provided in Table 9 of
the basic EMC standard
Conducted Disturbances Induced by RF
Fields (Conducted RF Immunity)
IEC 60601-1-2 / IEC 61000-4-6
3 Vrms
150 KHz - 80 MHz
6 Vrms
in ISM/amateur radio bands between
150 KHz and 80 MHz