Installation and Assembly
4-12 User Manual
DMR V4C 3.3 (User Manual) V-4 06-May-2022
Note:
Disconnecting this cable ceases power delivery to the V4C-O2E Oxygen Mixer.
4. Disconnect the gas connection tubing and the gas tube adapter from the ventilator.
Caution:
Ensure the appropriate filter is re-installed on the back of the ventilator's air intake before
initiating use of the ventilator without the V4C-O2E Oxygen Mixer. For more information on
filter type and assembly consult the manual of the compatible ventilator.
Caution:
Ensure the gas tube adapter is not left on the ventilator when the Oxygen Mixer is not being
used.
Note:
The compatible ventilator can be used without removing it from the Oxygen Mixer's handles. To use
it in this state see Section 5.4, Using the Compatible Ventilator Without the V4C-O2E Oxygen Mixer.
5. Using the installation tool, unscrew the two M6X25 bolts from the handle (bolt locations are shown in
Figure 4-6). Remove the handle from the sidearms.
6. Slide the compatible ventilator up and out of the Oxygen Mixer's side-arms.