Operating the bbe BenthoTorch
bbe BenthoTorch Manual Page 30
The algae have to be applied in 6 calibration plates. The volume of the cavity in the calibration plate is 100µl.
The area of the cavity is 1 cm². Please refer to the following table to get the required concentration of the algae
solution for application.
Number of the plate Concentration plate Concentration solution
Plate 1 1 µg/cm² diatoms 10000 µg/l
Plate 2 6 µg/cm² diatoms 60000 µg/l
Plate 3 1 µg/cm² cyanobacteria 10000 µg/l
Plate 4 6 µg/cm² cyanobacteria 60000 µg/l
Plate 5 1 µg/cm² green algae 10000 µg/l
Plate 6 6 µg/cm² green algae 60000 µg/l
To get a increase the concentration of the algae solution it can be centrifuged with about 2000 rpm for about
20min.
1. Apply 100µl of the first calibration solution to the first calibration plate
2. Put the BenthoTorch on the calibration plate