10
Miditherm MP
Basic principles
Up to 9 programs can be stored with Miditherm
(#1 to #9).
Up to 4 temperature holding levels can be set in each
program ( 1 to 4).
The following must be set at each holding level:
- heating-up speed (0 to 9°C/min.),
- temperature of the holding level (max. 1150°C),
- duration of the holding level (max. 99 min.).
The cut-out temperature of the connected Regulus
furnace extraction system can be set for each program.
The factory setting (600°C) is adequate for customary
investment materials that are heated up slowly. For
shock heat investment materials, the Regulus furnace
extraction system must be set to continuous operation
(see page 12).
Five standard programs are factory-set for BEGO al-
loys. The programs can be overwritten as required.
Your own programs should be noted. The tables in this
documentation (please make copies) serve this pur-
pose.
Notes
• In programs with fewer than 4 temperature
holding levels set all values of the following
holding levels to “0”.
• The furnace heats up at maximum speed when
the heating-up speed is set to “0°C/min.”.
• In seldom cases of lower power supply and
above 800 °C, the heating-up speed may not be
achieved.