13
Miditherm MP
Checking a program
Switch on unit. The current temperature of the
furnace is displayed after approx. 3 seconds.
Select program:
Press as often until LED lights up.
Press / until desired
program number appears.
ƒ Check values:
Press / to check the values of
heating-up speed (°C/min),
holding temperature (°C) and
holding time (min.).
To check the next holding level, go to the field
using the / buttons and call up the next
holding level using the / buttons.
Check the settings at this holding level as
described in point ƒ.
The current furnace temperature is displayed
again approx. 10 seconds after the last button was
pressed. The check is now finished.
3 x
22
22
2
Notes
The current furnace temperature is displayed
again approx. 10 seconds after the last button
was pressed. This also happens when hesitating
too long before pressing the next key.
Always check all holding levels! All values in the
unused holding levels of a program must be set
to “0”.
Changing values: see “Creating a new program”.