EasyManuals Logo

Bego Miditherm 100MP User Manual

Bego Miditherm 100MP
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #12 background imageLoading...
Page #12 background image
12
Miditherm MP
Creating a new programfor shock heat investment materials
So-called shock heat investment materials are not
heated up slowly, but are placed in a furnace that has
been preheated to the final temperature. As a result,
the modelling material burns atypically, i.e., if the
furnace door is opened, inflowing oxygen can cause a
darting flame!
Place all moulds in the furnace within 10
seconds; then keep the furnace door closed
for 15 minutes!
The cut-out of the Regulus furnace extraction system
is set to 600°C at the factory. Shock heat investment
materials are placed in the furnace at a higher
temperature and produce concentrated smoke gases
in a short space of time.
For shock heat investment materials, the
Regulus furnace extraction system (Ref.
25750) must be installed.
Regulus must be set to
continuous operation (see below).
to ƒ see “Creating a new program”
Call up and set the parameters at holding level 1:
heating-up speed: 0°C/min. (= max.)
temperature of the holding level:
see investment material
duration of the holding level:
see investment material
Set the parameters of holding levels 2, 3 and 4
to zero.
Press to set the extraction system to
continuous operation. The display shows “SH”
(shock heat).
(Press again to deactivate.)
Exit programming mode:
keep depressed until the LED “°C” lights up.
SHOCK HEAT
0 to 15 mins
START
STOP
SHOCK-
HEAT
#