5
Miditherm MP
Special safety instructions regarding this unit
l Installation, site of installation and start-up
Install unit only on adequately stable and non-combustible surface.
Because of the wax vapours, install the furnace under a chimney stalk as far as possible. Always
discharge vapours outdoors.
Only use the furnace in dry rooms.
There must not be any combustible objects in the area near the furnace.
There must not be any highly flammable or explosive gas, liquids or solid material in the same room as
the furnace.
Always connect the furnace to a separate circuit with 16-ampere fuse protection.
The unit is subject to safety class I and may only be connected to properly grounded power sources.
Use the rear plug only for connecting the Regulus furnace suction extraction unit (200-240 V).
Remove transport protection device from the mould chamber prior to start-up!
l Risk of burns due to hot parts
Wear protective clothing (safety apron), safety shoes, safety goggles and safety gloves.
Only touch the hot furnace door at the handle.
Move hot moulds only with sufficiently long tongs.
Let the unit cool down until it is warm to the touch before carrying out any care, maintenance or
cleaning work.
l Shock heat investment materials
So-called shock heat investment materials are not heated up slowly, but are placed in a furnace that has
been preheated to the final temperature. As a result, the modelling material burns atypically, i.e., if the
furnace door is opened, inflowing oxygen can cause a darting flame!
Place all moulds in the furnace within 10 seconds - then keep the furnace door closed for 15 minutes!
For shock heat investment materials, the Regulus furnace extraction system (Ref. 25750) must be
installed.
Regulus must be set to continuous operation (see page 12).
l Servicing work
Parts of the insulation contain carcinogenic ceramic fibres that may be released during servicing work.
Use gas mask type FFP2 – EN 149:2001 for servicing work with dust generation.