Maintenance
Any opening of the housing that is not described in
these operating instructions must only be carried
out by Customer Service staff or persons desig-
nated by Customer Service!
Metallic housings must be properly grounded to
prevent them from carrying electric current. If not
properly grounded, the housing may become live in
case of a damage inside of the device and thus may
pose a danger to life.
Only trained and qualified electricians must be
allowed to open the device as the device must be
checked for the absence of voltage at exposed,
conductive parts* after it has been opened.
* Germany: Testing in accordance with DIN VDE
0701-1.
Maintenance of safety-relevant compo-
nents
The device has been designed for a service life of
10 years from the date of manufacture. No liability is
accepted for damages arising from operation of the
device after this period.
Safety-relevant components must be checked
regularly and replaced as required. This work must
be carried out solely by BEGO customer service
personnel or by representatives that have been
authorized by the BEGO customer service
department. For that purpose, regular maintenance,
including annual inspections and an inspection after
five years, carried out by personnel authorized by
BEGO is recommended.
Regular maintenance
Regular maintenance is required to ensure proper
and faultless operation.
See see page 36 to display the total number of
casts.
Every 100 casts:
Clean the crucible and mould chamber (vacu-
um clean)
Clean the sight glass (see page 73),
Clean the seals (see page 74).
Every 500 casts:
Drain the pressure reducer (see page 75).
Every 3.000 casts:
Clean the ejector (see page 76),
Replace the air filter (see page 77).
Every 8.000 casts (or annually):
Inspection of safety-relevant components by
BEGO Customer Service
After 10.000 casts (or after 5 years):
Replacement of safety-relevant components by
BEGO Customer Service
DANGER
Warning of electric shock! Danger
to life!
Switch the device off and disconnect it
from the power supply before carrying
out any service or maintenance work!
(for power connection without mains
plug: remove fuses, secure the
devices against restart and make sure
that it is not live)
WARNING
Burn hazard from hot parts!
Carry out maintenance and cleaning
work before casting or after the device
has cooled down.

Other manuals for Bego Nautilus T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bego Nautilus T and is the answer not in the manual?

Bego Nautilus T Specifications

General IconGeneral
BrandBego
ModelNautilus T
CategoryDental equipment
LanguageEnglish

Related product manuals