LED surgical lights, single and combination lights CHROMOPHARE
®
E 778, E 668, E 558
Maintenance and inspection
85
– Due to possible damage to the materials, no
solutions are suitable that are based on
halogen-splitting bonds, strong organic acids
and oxygen-splitting bonds, solvents, fuel or
other.
– In order to avoid damages to the stainless steel
parts, only use disinfectants without chlorides or
halogenoids.
– Avoid contact between aldehyde and amino
products.
Therefore, especially if an amino-based
product was used previously, before applying
the aldehyde-based disinfectant for the first
time, an intermediate cleaning step has to be
performed. This is very important because
disregarding this step may result in residue no
longer being able to be removed.
– With regard to the microbiological
effectiveness, please contact the
manufacturer of the
disinfectant