What to do if my Binder ED 115 chamber doesn’t heat up and there is no heating icon in the display, but controller display is working?
- SSusan WilliamsAug 31, 2025
The timer may have run off. Try re-programming the timer or turn it off.
What to do if my Binder ED 115 chamber doesn’t heat up and there is no heating icon in the display, but controller display is working?
The timer may have run off. Try re-programming the timer or turn it off.
What to do if my ED, FD, FED Binder ED 115 Laboratory Equipment chamber has no function and shows alarm message “!TLim”?
This alarm indicates that safety device class 2 has turned off the chamber. Allow the chamber to cool down. Check the settings of the temperature set-point and the safety device class 2, and select a suitable limit value if needed.
What to do if my Binder Laboratory Equipment chamber has no function?
First, check the connection to the power supply. If the overtemperature protective device class 1 has turned off the chamber, disconnect the chamber from the power supply for at least 10 seconds and let it cool down. If the device responds again, further assistance is needed.
What to do if my ED, FD, or FED Binder ED 115 Laboratory Equipment chamber has no function and shows the alarm message “!TLim”?
The safety device class 2 has likely turned off the chamber. Allow the chamber to cool down. Check the temperature set-point and the settings of the safety device class 2. If appropriate, select a suitable limit value. If the issue persists, further assistance is needed.
Why is my Binder Laboratory Equipment chamber heating permanently and not holding the set-point?
The controller may not be adjusted, so calibrate and adjust it. If the issue persists, the controller might be defective and further assistance is needed.
What to do if Binder Laboratory Equipment chamber doesn’t heat up but the controller display is working?
The timer might have run off. Try re-programming the timer or turning it off.
Why is the set-point temperature not reached in my Binder Laboratory Equipment after the specified time?
Several factors could be the reason. Make sure the chamber door is completely closed and that the door gasket isn't defective (replace if needed). The controller may also need calibration and adjustment. Finally, verify the power supply for the correct voltage.
Qualification requirements for personnel operating the chamber.
Important safety guidelines for installation and operation.
Description of the chamber's proper applications and limitations.
Instructions for safely moving and transporting the chamber.
Requirements for the installation site and environmental conditions.
Detailed instructions and data for connecting the chamber electrically.
Description of the controller buttons, symbols, and display.
How to set temperature and fan speed values.
Procedure for setting user and controller access passwords.
Description of the class 1 overtemperature protection device.
Details on the adjustable electronic safety controller and its classes.
How to set the limit temperature for the safety controller.
How to handle alarm messages and system responses.
Overview of available timer functions and their operation.
Detailed steps for setting the "Delayed Off" timer function.
Steps for setting the temperature-dependent delayed off timer.
Setting timer run-time, fan, and set-point for temperature-dependent timer.
Steps for setting the "Delayed On" timer function.
Setting timer run-time and fan speed for Delayed On timer.
How to program and view ramp function settings.
Viewing and modifying the ramp set-point.
Instructions on how to initiate data recording via USB.
Instructions on how to stop data recording via USB.
How to manually set the chamber's IP address.
Procedures and recommended agents for cleaning the chamber.
Methods for decontaminating the chamber after exposure to hazardous substances.
Overview of maintenance, service, and personnel requirements.
Recommended intervals and procedures for maintenance.
Common faults and their solutions for operating personnel.
Description of overcurrent protection mechanisms.
Technical specifications for BD series chambers.
| Brand | Binder |
|---|---|
| Model | ED 115 |
| Category | Laboratory Equipment |
| Language | English |
Loading...